Hydrogen Peroxide Biosensors Based on the Immobilization of Hemoglobin on Amino-Silane Magnetic Nanoparticle-Contained Carbon Paste Electrodes

Huang, Y-M, Zhang, H-L, Yu, A-M, Liu, Y. and Wu, Y-W (2010) Hydrogen Peroxide Biosensors Based on the Immobilization of Hemoglobin on Amino-Silane Magnetic Nanoparticle-Contained Carbon Paste Electrodes. Sensor Letters, 8 (6). pp. 864-870.

Link to Published Version: http://dx.doi.org/10.1166/sl.2010.1357
*Subscription may be required

Abstract

Amino-silane magnetic nanoparticles (A-SMNPs) were mixed with graphite powder to construct A-SMNPs-contained carbon paste electrodes (CPE). Hemoglobin (Hb) was then immobilized on the electrode surface to form Hb/A-SMNPs/CPE biosensors. The biosensor was characterized by electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV). The results indicated that Hb was stably immobilized on the electrode and exhibit excellent electrocatalytic activity to the reduction of hydrogen peroxide without the aid of an electron mediator. When used for amperometric measurement of H2O2, the prepared biosensor displayed good sensitivity, selectivity and stability. A linear detection range of H2O2 from 1.0 mu mol L-1 to 2.0 mmol L-1 and a detection limit of 0.27 mu mol L-1 (S/N = 3) was obtained. The apparent Michaelis-Menten constant was calculated to be 0.85 mmol L-1, indicating that the immobilized Hb had high enzymatic activity to H2O2.

Publication Type: Journal Article
Murdoch Affiliation: School of Chemical and Mathematical Science
Publisher: American Scientific Publishers
Copyright: (c) American Scientific Publishers
URI: http://researchrepository.murdoch.edu.au/id/eprint/3477
Item Control Page