The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study

Golriz, S., Hebert, J.J., Foreman, K.B. and Walker, B.F. (2012) The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study. Chiropractic & Manual Therapies, 20 (1). p. 15.

[img]
Preview
PDF - Published Version
Download (544kB) | Preview
    Link to Published Version: http://dx.doi.org/10.1186/2045-709X-20-15
    *Open access, no subscription required

    Abstract

    Background
    The broad use of force plates in clinical settings for postural control assessment suggests the need for instruments that are easy to use, affordable and readily available. In addition, these instruments of measurement should be reliable and valid as adequate reliability and validity are prerequisites to making correct inferences. The aim of this study was to examine the concurrent validity of postural control measures obtained with a clinical force plate.

    Methods
    Thirty-one healthy adults were recruited. Participants completed 1 set of 5 trials on each force plate. Postural control measures (centre of pressure [COP] average velocity and sway area) were collected and compared using the Midot Posture Scale Analyzer (clinical force plate) and the Accugait force plate (criterion measure). Intra class correlation coefficient (ICC), standard error of measurement , and paired t-tests were calculated and Bland-Altman plots were constructed to compare the force plates and assess consistency of measurement and agreement between them.

    Results
    The ICC values (ICC = 0.14-0.60) between the two force plates were lower than the acceptable value for both COP average velocity and sway area. There was significant difference (p > 0.05) in COP average velocity and sway area between the force plates. Examination of the plots revealed that there is less difference between the force plates in lower magnitudes of COP for average velocity and sway area however, the greater the average velocity and sway area, the greater the difference between the measures obtained from the two force plates.

    Conclusion
    Findings of this study showed poor concurrent validity of the clinical force plate. This clinical force plate cannot be a replacement for known reliable and valid force plates and consequently measures obtained from this force plate should be treated with caution especially in a clinical population.

    Publication Type: Journal Article
    Murdoch Affiliation: School of Chiropractic and Sports Science
    Publisher: BioMed Central
    Copyright: © 2012 Golriz et al.
    Notes: This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
    URI: http://researchrepository.murdoch.edu.au/id/eprint/10755
    Item Control Page

    Downloads

    Downloads per month over past year